Laboratory Solutions for Material Science
 
  
 
 

DCAT 21 
Dynamic Contact Angle Meter and Tensiometer

Description:
The DCAT 21 is used for the software-controlled measurement of 
  • the contact angle of powders and fiber bundles and the adsorption behavior of liquids on wettable materials according to the modified Washburn method
  • the dynamic contact angle and the surface energy of prismatic and cylindrical samples according to the Wilhelmy method
  • the static surface and interfacial tension of liquids
  • the critical micelle formation concentration of surfactants (CMC) according to the Wilhelmy Plate method, Lenard Frame method or the Du Noüy Ring method with automatic ring correction according to Huh & Mason, Zuidema & Waters and Harkins & Jordan
as well as for fully automated reproducibility measurements.

Equipment:

The DCAT 21 consists of the basic instrument with the following technical features:

  • High-precision electrodynamically compensated weighing system with automatic calibrating function
  • Software-controlled motorized height positioning of the sample vessel with variable speed
  • Integrated measuring and control electronics with digital thermometer
  • Graphical display for weighing data, temperature and other measuring statuses
  • Serial RS 232C interface with PC connecting cable
  • Illuminated sample chamber with connection for inert gas
  • Sample vessels of borosilicate glass available with diameters 50,70,100 mm
  • Pt100 temperature sensor for the installation in the temperature control unit
  • Du Noüy ring RG 11, made of platinum-iridium to DIN 53914
  • Wilhelmy plate PT 11, made of platinum-iridium to DIN 53914
It is prepared for the installation of a liquid temperature control unit with integrated magnetic stirrer for mounting either of the sample vessels TV 50 / TV 70 / TV 100.

Technical Data:
 
Measuring Range : 1 … 1000 mN/m; ± 0.001 mN/m
Measuring Rate : 25 weighing data/s (optionally 50 weighing data/s)
Weighing Range : 0,1 mg … 210 g; ± 0,01 mg
Liftng Speed : 0,7 µm/s500 mm/min
Position Resolution : 0,1 µm 
Maximum lifting range : 105 mm
Temperature range : -10° … 130° C
Temperature measurement : 2 Pt100 input channels with -60 … 450°C (Pt100 optional), 0.1 K resolution; 1/3 DIN IEC 751 (± 0.03 %), class B
Measuring methods : Modified Washburn method
Dynamic Wilhelmy method
Wilhelmy Plate method
Lenard Frame method
Du Noüy Ring method
Receptacle for sample vessel 50 mm diameter (optionally 70 mm and 100 mm)
Size of instrument (L x W x H) 305 x 230 x 575 mm
Weight 25 kg
Power supply 90 … 264 VAC; 43 … 60 Hz; 55 VA
Alternative/Supplementary units and accessories : Liquid temperature control units (-10 …130°C); sample vessels; cylindrical and Wilhelmy plates; PTFE-sample bodies; Du Noüy rings; precision dosing system; software SCAT; complete PC-systems


© LMS Group of Companies 1995 - 2008. All Rights Reserved.
Singapore        ¤        Malaysia        ¤        Thailand        ¤        Indonesia        ¤         Vietnam