Laser-Particle-Sizer
"analysette 22" MicroTec plus
Wet Dispersing Unit
Description
The ANALYSETTE 22 MicroTec plus fascinates with an unachieved
union of extra-wide measuring range, high resolution and
sensitivity, marginal space requirements and a reasonable price.
The modular built ANALYSETTE MicroTec plus consists of a central
measuring unit for the angular aperture measuring of the
scattered light intensity and different dispersion modules,
which can be exchanged quickly and simply without tools. Hereby
it is without any problems possible to adjust the entire system
at a later date to possibly new requirements of the dispersion.
The particle size area covered by the measuring unit reaches
from 0.08 µm (80nm) up to 2000 µm.
The typical areas of application
The Laser particle measuring instrument ANALYSETTE 22 MicroTec
plus, is an universally applicable instrument for the
determination of the particle size distribution of samples in
powder form, solids in suspensions and of emulsions.
The instrument is
suitable for the utilization in quality and process control. Due
to the highly flexible programmability of the entire dispersion
and measuring process, it is also very apt for research and
development.
The Technology
The
larger measuring range is achieved by using two lasers with
different wave lengths. A green, linear polarized 7mW-Laser with
521nm wavelength is used in the area of fine particles, whereas
a second fibre-coupled IR-laser (Wavelength: 940nm) with also
7mW performance, is used for measuring larger grains.
Additionally, the
angular range of the diffracted respectively scattered laser
light, can be adjusted to the detector through the for FRITSCH
typical movable measuring cell.
With large
particles a large distance between the measuring cell and the
detector is selected, so that the here typical small diffraction
angles may still be detected very well. With fine and very fine
powders, a small measuring cell detector-distance should be
selected and obtains by this, optimal conditions for the
measuring of large scattering angles. This is the trick which
leads to a high resolution and an exceptional sensitivity.
By the means of a
folded optical path, a compact design is achieved and by the
adept arrangement of the measuring cell, a fast and safe switch
of the selectable measuring ranges.
The as a cassette
developed measuring cell allows the simple, tool-free exchange
between two different dispersion units, as well as the
comfortable cleaning of the measuring cell.
Features overview
- Extra large measuring
range 80nm to 2mm
- Up to 108 particle
size intervals
- Exceptional
price-performance-ratio
- Compact dimensions
- Fully automated
measuring course
- High reproducibility
- Automatic beam
adjustment
- Automatic positioning
of the measuring cell
- Measuring cell
Fast-Switch-System
- Adjustment of the
measuring range without modifications
- Detector on solid
ceramic support
- Interface for
individual programming
- 521nm laser diode for
fine particles
- 940nm fibre-coupled
laser diode for coarse particles
- Laser safety class 1
- Can be calibrated
according to ISO 9001:2000
- Meets the ISO 13320-1
(Laser Diffraction) requirements
Technical Data
|
Analysis
method |
: |
Laser
diffraction / Laser scattering |
|
|
Optical
set-up |
: |
Inverse
Fourier with movable measuring cell |
|
|
Measurement
range for wet dispersion |
: |
0.08
- 2000 µm |
|
|
Dispersion |
: |
wet
and dry |
|
|
Measurement
time |
: |
Approx.
10 s |
|
|
Liquid
quantity in the sample circuit |
: |
Approx.
400 ml |
|
|
Whole
measurement cycle |
: |
Typical
2 min |
|
|
Maximal
number of measurement channels |
: |
108 |
|
|
Reproducibility
acc. to ISO 13320-1 |
: |
d50
≤ 2% |
|
|
Optical
calculation methods |
: |
Mie,
Fraunhofer |
|
|
Laser
wavelength |
: |
512nm
and 940 nm |
|
|
Voltage |
: |
230 V/1~ |
|
|
Frequency |
: |
50 - 60 Hz |
|
|
Dimensions
(W x D x H) |
: |
Table
unit: 53x62x35cm |
|
|
Weight |
: |
38kg |
|
|