Online Hazemeter Gamma12
Measurement according to ASTM 1003
standard
The functional principle of Gamma12 complies with the standard ASTM
1003. The measurement sample is passed across the opening in an
integrating sphere so that parallel light can pass through the sample into
the sphere. The intensity of the dispersed light can be measured using a
high-precision sensor by means of the dispersion of the transmitted light
beam. Light beams with dispersion angles exceeding a standardised value
cannot leave the sphere through the exit and are reflected by the sphere
surface. The intensity of the scattered light is measured after multiple
reflections within the integrating sphere.
The intensity of the total light transmitted can be determined by
measuring with the sphere exit closed. The haze value is calculated from
the ratio of the scattered light and the total transmitted light.
In addition to the haze, the transmittance of the sample is measured.
This is derived from the ratio of the total transmitted light and the
intensity of the illumination. The measurements are shown online on a
display as percentages and therefore provide an objective indication of
quality instead of a visual subjective assessment.
The measurement results can be documented and analysed with
high-performance software tools owing to the link via the interface
RS232.
Performance features
- Online measurement
The measurements are made while the
sample is moving. No stop of production required.
- Easy touch
Simple menu-guided desktop with coloured touch
display
- Online measurement presentation
The measurement results
are continuously updated on the display
- Multi-point calibration
Simple multi-point calibration
- Compensation measurement
Continuous reference beam
measurement to compensate for changes in the intensity of the
illumination
- Fault diagnosis
System monitoring by means of a continuous
self-test of the hardware components
- Light technology
Choppered light source, therefore no
effect from temperature and ambient light
- Signal processing
Use of modern LOCK-IN amplifier
technology
- Embedded technology
Integrated embedded PC for unit
control
- Remote control
Remote control via serial RS-232 interface
- Open database format
The measurement data can be converted
into all common data formats (Access, Excel, ...) because of the link to
the Film Analyser Software
- Quality assurance
The RS232 interface permits connection
to a printer, e.g. for printing out statistics, and is therefore the
ideal preparation for ISO 900
|